Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis
Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has becom...
Ausführliche Beschreibung
Autor*in: |
Neha Sharma [verfasserIn] Jokin Zubizarreta-Oteiza [verfasserIn] Céline Tourbier [verfasserIn] Florian M. Thieringer [verfasserIn] |
---|
Format: |
E-Artikel |
---|---|
Sprache: |
Englisch |
Erschienen: |
2023 |
---|
Schlagwörter: |
---|
Übergeordnetes Werk: |
In: Journal of Clinical Medicine - MDPI AG, 2013, 12(2023), 7, p 2495 |
---|---|
Übergeordnetes Werk: |
volume:12 ; year:2023 ; number:7, p 2495 |
Links: |
---|
DOI / URN: |
10.3390/jcm12072495 |
---|
Katalog-ID: |
DOAJ089368339 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | DOAJ089368339 | ||
003 | DE-627 | ||
005 | 20240413043051.0 | ||
007 | cr uuu---uuuuu | ||
008 | 230505s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.3390/jcm12072495 |2 doi | |
035 | |a (DE-627)DOAJ089368339 | ||
035 | |a (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 0 | |a Neha Sharma |e verfasserin |4 aut | |
245 | 1 | 0 | |a Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a Computermedien |b c |2 rdamedia | ||
338 | |a Online-Ressource |b cr |2 rdacarrier | ||
520 | |a Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. | ||
650 | 4 | |a accuracy | |
650 | 4 | |a alloplastic implant | |
650 | 4 | |a computer-assisted | |
650 | 4 | |a cranioplasty | |
650 | 4 | |a material extrusion | |
650 | 4 | |a patient-specific | |
653 | 0 | |a Medicine | |
653 | 0 | |a R | |
700 | 0 | |a Jokin Zubizarreta-Oteiza |e verfasserin |4 aut | |
700 | 0 | |a Céline Tourbier |e verfasserin |4 aut | |
700 | 0 | |a Florian M. Thieringer |e verfasserin |4 aut | |
773 | 0 | 8 | |i In |t Journal of Clinical Medicine |d MDPI AG, 2013 |g 12(2023), 7, p 2495 |w (DE-627)718632478 |w (DE-600)2662592-1 |x 20770383 |7 nnns |
773 | 1 | 8 | |g volume:12 |g year:2023 |g number:7, p 2495 |
856 | 4 | 0 | |u https://doi.org/10.3390/jcm12072495 |z kostenfrei |
856 | 4 | 0 | |u https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 |z kostenfrei |
856 | 4 | 0 | |u https://www.mdpi.com/2077-0383/12/7/2495 |z kostenfrei |
856 | 4 | 2 | |u https://doaj.org/toc/2077-0383 |y Journal toc |z kostenfrei |
912 | |a GBV_USEFLAG_A | ||
912 | |a SYSFLAG_A | ||
912 | |a GBV_DOAJ | ||
912 | |a GBV_ILN_20 | ||
912 | |a GBV_ILN_22 | ||
912 | |a GBV_ILN_23 | ||
912 | |a GBV_ILN_24 | ||
912 | |a GBV_ILN_39 | ||
912 | |a GBV_ILN_40 | ||
912 | |a GBV_ILN_60 | ||
912 | |a GBV_ILN_62 | ||
912 | |a GBV_ILN_63 | ||
912 | |a GBV_ILN_65 | ||
912 | |a GBV_ILN_69 | ||
912 | |a GBV_ILN_73 | ||
912 | |a GBV_ILN_74 | ||
912 | |a GBV_ILN_95 | ||
912 | |a GBV_ILN_105 | ||
912 | |a GBV_ILN_110 | ||
912 | |a GBV_ILN_151 | ||
912 | |a GBV_ILN_161 | ||
912 | |a GBV_ILN_170 | ||
912 | |a GBV_ILN_206 | ||
912 | |a GBV_ILN_213 | ||
912 | |a GBV_ILN_230 | ||
912 | |a GBV_ILN_285 | ||
912 | |a GBV_ILN_293 | ||
912 | |a GBV_ILN_602 | ||
912 | |a GBV_ILN_2005 | ||
912 | |a GBV_ILN_2009 | ||
912 | |a GBV_ILN_2014 | ||
912 | |a GBV_ILN_2055 | ||
912 | |a GBV_ILN_2111 | ||
912 | |a GBV_ILN_4012 | ||
912 | |a GBV_ILN_4037 | ||
912 | |a GBV_ILN_4112 | ||
912 | |a GBV_ILN_4125 | ||
912 | |a GBV_ILN_4126 | ||
912 | |a GBV_ILN_4249 | ||
912 | |a GBV_ILN_4305 | ||
912 | |a GBV_ILN_4306 | ||
912 | |a GBV_ILN_4307 | ||
912 | |a GBV_ILN_4313 | ||
912 | |a GBV_ILN_4322 | ||
912 | |a GBV_ILN_4323 | ||
912 | |a GBV_ILN_4324 | ||
912 | |a GBV_ILN_4325 | ||
912 | |a GBV_ILN_4338 | ||
912 | |a GBV_ILN_4367 | ||
912 | |a GBV_ILN_4700 | ||
951 | |a AR | ||
952 | |d 12 |j 2023 |e 7, p 2495 |
author_variant |
n s ns j z o jzo c t ct f m t fmt |
---|---|
matchkey_str |
article:20770383:2023----::asemtrlztoafctecuayfonocr3pitdoyteehreoeekutmzdrn |
hierarchy_sort_str |
2023 |
publishDate |
2023 |
allfields |
10.3390/jcm12072495 doi (DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 DE-627 ger DE-627 rakwb eng Neha Sharma verfasserin aut Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis 2023 Text txt rdacontent Computermedien c rdamedia Online-Ressource cr rdacarrier Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R Jokin Zubizarreta-Oteiza verfasserin aut Céline Tourbier verfasserin aut Florian M. Thieringer verfasserin aut In Journal of Clinical Medicine MDPI AG, 2013 12(2023), 7, p 2495 (DE-627)718632478 (DE-600)2662592-1 20770383 nnns volume:12 year:2023 number:7, p 2495 https://doi.org/10.3390/jcm12072495 kostenfrei https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 kostenfrei https://www.mdpi.com/2077-0383/12/7/2495 kostenfrei https://doaj.org/toc/2077-0383 Journal toc kostenfrei GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 AR 12 2023 7, p 2495 |
spelling |
10.3390/jcm12072495 doi (DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 DE-627 ger DE-627 rakwb eng Neha Sharma verfasserin aut Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis 2023 Text txt rdacontent Computermedien c rdamedia Online-Ressource cr rdacarrier Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R Jokin Zubizarreta-Oteiza verfasserin aut Céline Tourbier verfasserin aut Florian M. Thieringer verfasserin aut In Journal of Clinical Medicine MDPI AG, 2013 12(2023), 7, p 2495 (DE-627)718632478 (DE-600)2662592-1 20770383 nnns volume:12 year:2023 number:7, p 2495 https://doi.org/10.3390/jcm12072495 kostenfrei https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 kostenfrei https://www.mdpi.com/2077-0383/12/7/2495 kostenfrei https://doaj.org/toc/2077-0383 Journal toc kostenfrei GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 AR 12 2023 7, p 2495 |
allfields_unstemmed |
10.3390/jcm12072495 doi (DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 DE-627 ger DE-627 rakwb eng Neha Sharma verfasserin aut Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis 2023 Text txt rdacontent Computermedien c rdamedia Online-Ressource cr rdacarrier Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R Jokin Zubizarreta-Oteiza verfasserin aut Céline Tourbier verfasserin aut Florian M. Thieringer verfasserin aut In Journal of Clinical Medicine MDPI AG, 2013 12(2023), 7, p 2495 (DE-627)718632478 (DE-600)2662592-1 20770383 nnns volume:12 year:2023 number:7, p 2495 https://doi.org/10.3390/jcm12072495 kostenfrei https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 kostenfrei https://www.mdpi.com/2077-0383/12/7/2495 kostenfrei https://doaj.org/toc/2077-0383 Journal toc kostenfrei GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 AR 12 2023 7, p 2495 |
allfieldsGer |
10.3390/jcm12072495 doi (DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 DE-627 ger DE-627 rakwb eng Neha Sharma verfasserin aut Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis 2023 Text txt rdacontent Computermedien c rdamedia Online-Ressource cr rdacarrier Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R Jokin Zubizarreta-Oteiza verfasserin aut Céline Tourbier verfasserin aut Florian M. Thieringer verfasserin aut In Journal of Clinical Medicine MDPI AG, 2013 12(2023), 7, p 2495 (DE-627)718632478 (DE-600)2662592-1 20770383 nnns volume:12 year:2023 number:7, p 2495 https://doi.org/10.3390/jcm12072495 kostenfrei https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 kostenfrei https://www.mdpi.com/2077-0383/12/7/2495 kostenfrei https://doaj.org/toc/2077-0383 Journal toc kostenfrei GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 AR 12 2023 7, p 2495 |
allfieldsSound |
10.3390/jcm12072495 doi (DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 DE-627 ger DE-627 rakwb eng Neha Sharma verfasserin aut Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis 2023 Text txt rdacontent Computermedien c rdamedia Online-Ressource cr rdacarrier Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R Jokin Zubizarreta-Oteiza verfasserin aut Céline Tourbier verfasserin aut Florian M. Thieringer verfasserin aut In Journal of Clinical Medicine MDPI AG, 2013 12(2023), 7, p 2495 (DE-627)718632478 (DE-600)2662592-1 20770383 nnns volume:12 year:2023 number:7, p 2495 https://doi.org/10.3390/jcm12072495 kostenfrei https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 kostenfrei https://www.mdpi.com/2077-0383/12/7/2495 kostenfrei https://doaj.org/toc/2077-0383 Journal toc kostenfrei GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 AR 12 2023 7, p 2495 |
language |
English |
source |
In Journal of Clinical Medicine 12(2023), 7, p 2495 volume:12 year:2023 number:7, p 2495 |
sourceStr |
In Journal of Clinical Medicine 12(2023), 7, p 2495 volume:12 year:2023 number:7, p 2495 |
format_phy_str_mv |
Article |
institution |
findex.gbv.de |
topic_facet |
accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific Medicine R |
isfreeaccess_bool |
true |
container_title |
Journal of Clinical Medicine |
authorswithroles_txt_mv |
Neha Sharma @@aut@@ Jokin Zubizarreta-Oteiza @@aut@@ Céline Tourbier @@aut@@ Florian M. Thieringer @@aut@@ |
publishDateDaySort_date |
2023-01-01T00:00:00Z |
hierarchy_top_id |
718632478 |
id |
DOAJ089368339 |
language_de |
englisch |
fullrecord |
<?xml version="1.0" encoding="UTF-8"?><collection xmlns="http://www.loc.gov/MARC21/slim"><record><leader>01000caa a22002652 4500</leader><controlfield tag="001">DOAJ089368339</controlfield><controlfield tag="003">DE-627</controlfield><controlfield tag="005">20240413043051.0</controlfield><controlfield tag="007">cr uuu---uuuuu</controlfield><controlfield tag="008">230505s2023 xx |||||o 00| ||eng c</controlfield><datafield tag="024" ind1="7" ind2=" "><subfield code="a">10.3390/jcm12072495</subfield><subfield code="2">doi</subfield></datafield><datafield tag="035" ind1=" " ind2=" "><subfield code="a">(DE-627)DOAJ089368339</subfield></datafield><datafield tag="035" ind1=" " ind2=" "><subfield code="a">(DE-599)DOAJ7eab7784986b471b9c665ee52ad57265</subfield></datafield><datafield tag="040" ind1=" " ind2=" "><subfield code="a">DE-627</subfield><subfield code="b">ger</subfield><subfield code="c">DE-627</subfield><subfield code="e">rakwb</subfield></datafield><datafield tag="041" ind1=" " ind2=" "><subfield code="a">eng</subfield></datafield><datafield tag="100" ind1="0" ind2=" "><subfield code="a">Neha Sharma</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="245" ind1="1" ind2="0"><subfield code="a">Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis</subfield></datafield><datafield tag="264" ind1=" " ind2="1"><subfield code="c">2023</subfield></datafield><datafield tag="336" ind1=" " ind2=" "><subfield code="a">Text</subfield><subfield code="b">txt</subfield><subfield code="2">rdacontent</subfield></datafield><datafield tag="337" ind1=" " ind2=" "><subfield code="a">Computermedien</subfield><subfield code="b">c</subfield><subfield code="2">rdamedia</subfield></datafield><datafield tag="338" ind1=" " ind2=" "><subfield code="a">Online-Ressource</subfield><subfield code="b">cr</subfield><subfield code="2">rdacarrier</subfield></datafield><datafield tag="520" ind1=" " ind2=" "><subfield code="a">Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants.</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">accuracy</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">alloplastic implant</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">computer-assisted</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">cranioplasty</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">material extrusion</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">patient-specific</subfield></datafield><datafield tag="653" ind1=" " ind2="0"><subfield code="a">Medicine</subfield></datafield><datafield tag="653" ind1=" " ind2="0"><subfield code="a">R</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Jokin Zubizarreta-Oteiza</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Céline Tourbier</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Florian M. Thieringer</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="773" ind1="0" ind2="8"><subfield code="i">In</subfield><subfield code="t">Journal of Clinical Medicine</subfield><subfield code="d">MDPI AG, 2013</subfield><subfield code="g">12(2023), 7, p 2495</subfield><subfield code="w">(DE-627)718632478</subfield><subfield code="w">(DE-600)2662592-1</subfield><subfield code="x">20770383</subfield><subfield code="7">nnns</subfield></datafield><datafield tag="773" ind1="1" ind2="8"><subfield code="g">volume:12</subfield><subfield code="g">year:2023</subfield><subfield code="g">number:7, p 2495</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://doi.org/10.3390/jcm12072495</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://doaj.org/article/7eab7784986b471b9c665ee52ad57265</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://www.mdpi.com/2077-0383/12/7/2495</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="2"><subfield code="u">https://doaj.org/toc/2077-0383</subfield><subfield code="y">Journal toc</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_USEFLAG_A</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">SYSFLAG_A</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_DOAJ</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_20</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_22</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_23</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_24</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_39</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_40</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_60</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_62</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_63</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_65</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_69</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_73</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_74</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_95</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_105</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_110</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_151</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_161</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_170</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_206</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_213</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_230</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_285</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_293</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_602</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2005</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2009</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2014</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2055</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2111</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4012</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4037</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4112</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4125</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4126</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4249</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4305</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4306</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4307</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4313</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4322</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4323</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4324</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4325</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4338</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4367</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4700</subfield></datafield><datafield tag="951" ind1=" " ind2=" "><subfield code="a">AR</subfield></datafield><datafield tag="952" ind1=" " ind2=" "><subfield code="d">12</subfield><subfield code="j">2023</subfield><subfield code="e">7, p 2495</subfield></datafield></record></collection>
|
author |
Neha Sharma |
spellingShingle |
Neha Sharma misc accuracy misc alloplastic implant misc computer-assisted misc cranioplasty misc material extrusion misc patient-specific misc Medicine misc R Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
authorStr |
Neha Sharma |
ppnlink_with_tag_str_mv |
@@773@@(DE-627)718632478 |
format |
electronic Article |
delete_txt_mv |
keep |
author_role |
aut aut aut aut |
collection |
DOAJ |
remote_str |
true |
illustrated |
Not Illustrated |
issn |
20770383 |
topic_title |
Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis accuracy alloplastic implant computer-assisted cranioplasty material extrusion patient-specific |
topic |
misc accuracy misc alloplastic implant misc computer-assisted misc cranioplasty misc material extrusion misc patient-specific misc Medicine misc R |
topic_unstemmed |
misc accuracy misc alloplastic implant misc computer-assisted misc cranioplasty misc material extrusion misc patient-specific misc Medicine misc R |
topic_browse |
misc accuracy misc alloplastic implant misc computer-assisted misc cranioplasty misc material extrusion misc patient-specific misc Medicine misc R |
format_facet |
Elektronische Aufsätze Aufsätze Elektronische Ressource |
format_main_str_mv |
Text Zeitschrift/Artikel |
carriertype_str_mv |
cr |
hierarchy_parent_title |
Journal of Clinical Medicine |
hierarchy_parent_id |
718632478 |
hierarchy_top_title |
Journal of Clinical Medicine |
isfreeaccess_txt |
true |
familylinks_str_mv |
(DE-627)718632478 (DE-600)2662592-1 |
title |
Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
ctrlnum |
(DE-627)DOAJ089368339 (DE-599)DOAJ7eab7784986b471b9c665ee52ad57265 |
title_full |
Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
author_sort |
Neha Sharma |
journal |
Journal of Clinical Medicine |
journalStr |
Journal of Clinical Medicine |
lang_code |
eng |
isOA_bool |
true |
recordtype |
marc |
publishDateSort |
2023 |
contenttype_str_mv |
txt |
author_browse |
Neha Sharma Jokin Zubizarreta-Oteiza Céline Tourbier Florian M. Thieringer |
container_volume |
12 |
format_se |
Elektronische Aufsätze |
author-letter |
Neha Sharma |
doi_str_mv |
10.3390/jcm12072495 |
author2-role |
verfasserin |
title_sort |
can steam sterilization affect the accuracy of point-of-care 3d printed polyetheretherketone (peek) customized cranial implants? an investigative analysis |
title_auth |
Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
abstract |
Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. |
abstractGer |
Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. |
abstract_unstemmed |
Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants. |
collection_details |
GBV_USEFLAG_A SYSFLAG_A GBV_DOAJ GBV_ILN_20 GBV_ILN_22 GBV_ILN_23 GBV_ILN_24 GBV_ILN_39 GBV_ILN_40 GBV_ILN_60 GBV_ILN_62 GBV_ILN_63 GBV_ILN_65 GBV_ILN_69 GBV_ILN_73 GBV_ILN_74 GBV_ILN_95 GBV_ILN_105 GBV_ILN_110 GBV_ILN_151 GBV_ILN_161 GBV_ILN_170 GBV_ILN_206 GBV_ILN_213 GBV_ILN_230 GBV_ILN_285 GBV_ILN_293 GBV_ILN_602 GBV_ILN_2005 GBV_ILN_2009 GBV_ILN_2014 GBV_ILN_2055 GBV_ILN_2111 GBV_ILN_4012 GBV_ILN_4037 GBV_ILN_4112 GBV_ILN_4125 GBV_ILN_4126 GBV_ILN_4249 GBV_ILN_4305 GBV_ILN_4306 GBV_ILN_4307 GBV_ILN_4313 GBV_ILN_4322 GBV_ILN_4323 GBV_ILN_4324 GBV_ILN_4325 GBV_ILN_4338 GBV_ILN_4367 GBV_ILN_4700 |
container_issue |
7, p 2495 |
title_short |
Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis |
url |
https://doi.org/10.3390/jcm12072495 https://doaj.org/article/7eab7784986b471b9c665ee52ad57265 https://www.mdpi.com/2077-0383/12/7/2495 https://doaj.org/toc/2077-0383 |
remote_bool |
true |
author2 |
Jokin Zubizarreta-Oteiza Céline Tourbier Florian M. Thieringer |
author2Str |
Jokin Zubizarreta-Oteiza Céline Tourbier Florian M. Thieringer |
ppnlink |
718632478 |
mediatype_str_mv |
c |
isOA_txt |
true |
hochschulschrift_bool |
false |
doi_str |
10.3390/jcm12072495 |
up_date |
2024-07-03T22:43:15.775Z |
_version_ |
1803599579220803584 |
fullrecord_marcxml |
<?xml version="1.0" encoding="UTF-8"?><collection xmlns="http://www.loc.gov/MARC21/slim"><record><leader>01000caa a22002652 4500</leader><controlfield tag="001">DOAJ089368339</controlfield><controlfield tag="003">DE-627</controlfield><controlfield tag="005">20240413043051.0</controlfield><controlfield tag="007">cr uuu---uuuuu</controlfield><controlfield tag="008">230505s2023 xx |||||o 00| ||eng c</controlfield><datafield tag="024" ind1="7" ind2=" "><subfield code="a">10.3390/jcm12072495</subfield><subfield code="2">doi</subfield></datafield><datafield tag="035" ind1=" " ind2=" "><subfield code="a">(DE-627)DOAJ089368339</subfield></datafield><datafield tag="035" ind1=" " ind2=" "><subfield code="a">(DE-599)DOAJ7eab7784986b471b9c665ee52ad57265</subfield></datafield><datafield tag="040" ind1=" " ind2=" "><subfield code="a">DE-627</subfield><subfield code="b">ger</subfield><subfield code="c">DE-627</subfield><subfield code="e">rakwb</subfield></datafield><datafield tag="041" ind1=" " ind2=" "><subfield code="a">eng</subfield></datafield><datafield tag="100" ind1="0" ind2=" "><subfield code="a">Neha Sharma</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="245" ind1="1" ind2="0"><subfield code="a">Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis</subfield></datafield><datafield tag="264" ind1=" " ind2="1"><subfield code="c">2023</subfield></datafield><datafield tag="336" ind1=" " ind2=" "><subfield code="a">Text</subfield><subfield code="b">txt</subfield><subfield code="2">rdacontent</subfield></datafield><datafield tag="337" ind1=" " ind2=" "><subfield code="a">Computermedien</subfield><subfield code="b">c</subfield><subfield code="2">rdamedia</subfield></datafield><datafield tag="338" ind1=" " ind2=" "><subfield code="a">Online-Ressource</subfield><subfield code="b">cr</subfield><subfield code="2">rdacarrier</subfield></datafield><datafield tag="520" ind1=" " ind2=" "><subfield code="a">Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (<i<p</i< < 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants.</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">accuracy</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">alloplastic implant</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">computer-assisted</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">cranioplasty</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">material extrusion</subfield></datafield><datafield tag="650" ind1=" " ind2="4"><subfield code="a">patient-specific</subfield></datafield><datafield tag="653" ind1=" " ind2="0"><subfield code="a">Medicine</subfield></datafield><datafield tag="653" ind1=" " ind2="0"><subfield code="a">R</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Jokin Zubizarreta-Oteiza</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Céline Tourbier</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="700" ind1="0" ind2=" "><subfield code="a">Florian M. Thieringer</subfield><subfield code="e">verfasserin</subfield><subfield code="4">aut</subfield></datafield><datafield tag="773" ind1="0" ind2="8"><subfield code="i">In</subfield><subfield code="t">Journal of Clinical Medicine</subfield><subfield code="d">MDPI AG, 2013</subfield><subfield code="g">12(2023), 7, p 2495</subfield><subfield code="w">(DE-627)718632478</subfield><subfield code="w">(DE-600)2662592-1</subfield><subfield code="x">20770383</subfield><subfield code="7">nnns</subfield></datafield><datafield tag="773" ind1="1" ind2="8"><subfield code="g">volume:12</subfield><subfield code="g">year:2023</subfield><subfield code="g">number:7, p 2495</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://doi.org/10.3390/jcm12072495</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://doaj.org/article/7eab7784986b471b9c665ee52ad57265</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="0"><subfield code="u">https://www.mdpi.com/2077-0383/12/7/2495</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="856" ind1="4" ind2="2"><subfield code="u">https://doaj.org/toc/2077-0383</subfield><subfield code="y">Journal toc</subfield><subfield code="z">kostenfrei</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_USEFLAG_A</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">SYSFLAG_A</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_DOAJ</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_20</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_22</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_23</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_24</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_39</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_40</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_60</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_62</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_63</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_65</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_69</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_73</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_74</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_95</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_105</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_110</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_151</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_161</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_170</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_206</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_213</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_230</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_285</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_293</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_602</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2005</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2009</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2014</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2055</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_2111</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4012</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4037</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4112</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4125</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4126</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4249</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4305</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4306</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4307</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4313</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4322</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4323</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4324</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4325</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4338</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4367</subfield></datafield><datafield tag="912" ind1=" " ind2=" "><subfield code="a">GBV_ILN_4700</subfield></datafield><datafield tag="951" ind1=" " ind2=" "><subfield code="a">AR</subfield></datafield><datafield tag="952" ind1=" " ind2=" "><subfield code="d">12</subfield><subfield code="j">2023</subfield><subfield code="e">7, p 2495</subfield></datafield></record></collection>
|
score |
7.399088 |