Minimum requirements for commissioning and long-term quality assurance of Elekta multi-leaf collimator for volumetric modulated arc therapy

Abstract We have proposed minimum requirements for commissioning and long-term quality assurance (QA) of an Elekta multi-leaf collimator (MLC) for volumetric modulated arc therapy (VMAT). The MLC leaf position accuracy during VMAT delivery was evaluated with the use of three different QA test plans:...
Ausführliche Beschreibung

Gespeichert in:
Autor*in:

Tatsumi, Daisaku [verfasserIn]

Nakada, Ryosei

Yomoda, Akane

Ishii, Kentaro

Tsutsumi, Shinichi

Inoue, Makoto

Ichida, Takao

Hosono, Masako N.

Miki, Yukio

Format:

E-Artikel

Sprache:

Englisch

Erschienen:

2012

Schlagwörter:

Volumetric modulated arc therapy

Multi-leaf collimator

Commissioning

Quality assurance

Leaf position accuracy

Anmerkung:

© Japanese Society of Radiological Technology and Japan Society of Medical Physics 2012

Übergeordnetes Werk:

Enthalten in: Radiological physics and technology - Tokyo : Springer, 2008, 6(2012), 1 vom: 14. Aug., Seite 98-106

Übergeordnetes Werk:

volume:6 ; year:2012 ; number:1 ; day:14 ; month:08 ; pages:98-106

Links:

Volltext

DOI / URN:

10.1007/s12194-012-0175-0

Katalog-ID:

SPR025177206

Nicht das Richtige dabei?

Schreiben Sie uns!