Dynamic positioning accuracy of a novel multileaf collimator for volumetric modulated arc therapy

Abstract We investigated the dynamic positioning accuracy of Agility (Elekta) for volumetric modulated arc therapy (VMAT). The accuracy of the multileaf collimator (MLC) leaf position during VMAT was evaluated using three different tests: (1) a dynamic multileaf collimator (DMLC) output test with va...
Ausführliche Beschreibung

Gespeichert in:
Autor*in:

Nakaguchi, Yuji [verfasserIn]

Ono, Takeshi

Onizuka, Ryota

Maruyama, Masato

Shimohigashi, Yoshinobu

Kai, Yudai

Format:

E-Artikel

Sprache:

Englisch

Erschienen:

2015

Schlagwörter:

Volumetric modulated arc therapy

Multileaf collimator

Quality assurance

DMLC

Leaf position accuracy

Anmerkung:

© Japanese Society of Radiological Technology and Japan Society of Medical Physics 2015

Übergeordnetes Werk:

Enthalten in: Radiological physics and technology - Tokyo : Springer, 2008, 9(2015), 1 vom: 26. Nov., Seite 121-126